Page view 4
Chapter 5.Measurements and Calculations 5 - 5
Distance Measurement
2D Distance (Line) Measurement: measures the distance between two
points
The 2D Distance measurement function is available in all diagnosis modes. Its results are
shown in cm (or mm). A maximum of four sets of measurements can be displayed at one
time.
1. Press the Caliper button, and the measurement item appears on the left side of the
screen.
2. Select ‘2D Distance (Line)’ by pressing Flexible Soft button [1] Distance or by
pressing Change or Caliper button several times.
3. Use the Trackball to position the cursor on the start and end points of the area to be
measured. Press the Set button to fix the points.
4. Pressing the Change button before pressing Set button allows the measurement area
to be reset.
5. After position setup, distance and average are displayed on the screen.
6. A new cursor appears for a new measurement. If more than four distance
measurements are performed, the distance measurements are calculated in order.
The ratio of the first distance to the second, and the ratio of the third distance to the fourth
are calculated and displayed as percentages.
2D Distance (Trace) Measurement: measures a non-straight distance
between two points
This measurement function can be performed in all diagnosis modes. The results are
displayed in cm (or mm) on the screen. A maximum of four sets of measurements can be
displayed at one time.
1. Press the Caliper button, and the measurement item appears on the left side of the
screen.
2. Select ‘2D Distance (Trace)’ by pressing Flexible Soft button [1] Distance or by
pressing Change or Caliper button several times.
3. Use the Trackball to position the cursor on the start and end points. Press the Set
button to fix the points.
4. Pressing the Change button before pressing Set button allows the measurement
area to be reset. Pressing the Del on the keyboard before pressing Set button
erases part of the trace line.
5. After position setup, distance and average are displayed on the screen.
Page view 4