Samsung SONOACE R7 User Manual

Browse online or download User Manual for Measuring, testing & control Samsung SONOACE R7. Samsung SONOACE R7 User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 224
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews

Summary of Contents

Page 3

7-60Operation ManualTips!Add Reference EquationIf a reference appears in an equation, the following should be entered:EquationEnter a reference equati

Page 4

7-61Chapter 7 Utility  Calc & GraphSelect the sub tab Calc & Graph under the OB tab. You can specify settings for calculation and graphs.[Fi

Page 5 - Table of Contents – Volume 2

7-62Operation ManualCardiac Measurement SetupSelect the Cardiac tab in the Measure Setup screen. You can specify settings for cardiac measurement. 

Page 6

7-63Chapter 7 UtilityVascular Measurement SetupSelect the Vascular tab in the Measure Setup screen. You can specify settings for vascular measurement

Page 7 - ** Reference Manual

7-64Operation ManualUrology Measurement SetupSelect the Urology tab in the Measure Setup screen. You can specify settings for urology measurement. [Fi

Page 8

7-65Chapter 7 Utility  Volume MethodSpecify an equation that will be used for volume calculation.X 3 Distance: The volume is calculated using thre

Page 9 - Image Management

7-66Operation ManualFetal Echo Measurement SetupSelect the Fetal Echo tab in the Measure Setup screen. You can specify settings for fetal echo measure

Page 10

7-67Chapter 7 Utility UtilitySpecify the system settings or use ECG, Biopsy or Histogram.1. Press the Utility button on the control panel. The Utili

Page 11 - Chapter 6 Image Management

7-68Operation Manual  Soft MenuChange the application or the probe settings, or adjust the monitor brightness.Applications / App. LoadUse the Soft

Page 12 -  How to Review Images

7-69Chapter 7 UtilityECGSelect the ECG in the Utility menu. The ECG (echocardiogram) function allows viewing of the cardiac pulsation. It shows an im

Page 13 - Cine Save

6-3Chapter 6 Image Management Reviewing Images (Cine / Loop)Images are automatically saved in the memory during scanning. The saved images can be us

Page 14 - Annotating Images

7-70Operation Manual Position / LoopX Position: Use the Soft Menu dial-button [5] Position to set the position of ECG. When its number is large, EC

Page 15 -  Typing Text

7-71Chapter 7 UtilityBiopsyPress Biopsy in the Utility menu. CAUTION: Before performing biopsy, you should specify the Biopsy Guide Line.[Figure 7.40

Page 16 -  Exiting Text Input Mode

7-72Operation Manual  Editing Biopsy Guide LineNOTE: XNote that the Biopsy Guide Line cannot be edited when the Trapezoidal function is in use for t

Page 17 - BodyMarker

7-73Chapter 7 UtilityHistogramA histogram is a type of graph representing the distribution of echoes. 1. Press Histogram in the Utility menu. 2. Sp

Page 18 -  Deleting BodyMarker

7-74Operation Manual[Figure 7.41 Histogram]

Page 19 - Indicator

7-75Chapter 7 UtilityPost CurvePress Post Curve in the Utility menu. Specify various post map and Gamma values. [Figure 7.42 Post Curve]  2D PostSel

Page 20 -  Deleting Indicator

7-76Operation ManualWhen you select this, the following item is enabled in the Soft Menu.XSoft Menu dial-button [1] Picker Pos: Select a point on a c

Page 21

7-77Chapter 7 Utility TagRotate the Menu dial-button to turn it on/o. When turned on, a random color is applied to a certain area in an image. T

Page 22 -  e-Motion Marker Setting

7-78Operation Manual  M PostSelect M Post in the Post Curve menu. The settings can be specied in the same way as for 2D Post.NOTE: For the post curv

Page 23 -  Image Save/Review

7-79Chapter 7 Utility EditWith the Soft Menu dial-button, adjust the RGB Curve. NOTE: This option is activated only when Curve is set to User1 – 3.

Page 24 - Saving Images

6-4Operation Manual  How to Review ImagesUse the Trackball on the control panel. You can search for saved images in chronological order by rotating t

Page 25

7-80Operation ManualSetupPress Setup in the Utility menu. You can specify system settings. Please refer to “System Setup” in this chapter.Measure Setu

Page 26 - Playing Images

7-81Chapter 7 UtilityStorage ManagerPress the Storage Manager in the Utility menu. All disk drives mounted in the system will be shown. The drive typ

Page 27 - Transferring Images

7-82Operation Manual Eject / RemoveDisconnect the selected drive. NOTE: Before unplugging a USB Flash memory drive, make sure to disconnect it by us

Page 28 - Recording Images

7-83Chapter 7 UtilityBodyMarker EditSelect BodyMarker Edit in the Utility menu. The BodyMarker Edit window appears.12[Figure 7.49 BodyMarker Edit]1

Page 29 - SonoView

7-84Operation Manual Removing a BodyMarkerSelect and double-click a BodyMarker in the right list (2).Saving and Exiting the BodyMarker ListTo save

Page 30 -  Selecting Exam

7-85Chapter 7 Utilitye-Motion MarkerSelect this if you are using e-Motion Marker, which is an optional feature for this product. NOTE: Xe-Motion Mar

Page 31 - Compare Mode

7-86Operation ManualMenu EditSelect Menu Edit in the Utility menu. The Menu Edit window appears. You can set a desired Soft Menu dial-button. 1234 5 6

Page 32 - Managing Exam Images

7-87Chapter 7 Utility5 Default Setting: Press the Soft Menu dial-button [2] to return to the default setting. 6 Delete: When the dial-button menu

Page 33

7-88Operation Manual Stress Echo (Optional)The Stress Echo function is used when added burden is put on the heart via exercise or drugs, in order to

Page 34 -  Measuring Distance

7-89Chapter 7 UtilitySetting Up a ProtocolWith no patient ID registered, press the Stress Echo button in the Utility menu. Alternatively, click the T

Page 35 -  Printing Images

6-5Chapter 6 Image Management Trim Last PosUse dial button [5] on the soft menu. Set the last frame of the range for saving when saving Cine or Loo

Page 36

7-90Operation Manual ProtocolWhen you click the combo button, the default protocols and other existing protocols are listed. Tips!Default ProtocolsT

Page 37 -  Transferring Images

7-91Chapter 7 Utility Capture UnitSelect a method for image acquisition by pressing the Save button on the control panel. XBeat: Acquires images b

Page 38 -  Sending in E-mail

7-92Operation Manual ExportPressing the Export button opens the Protocol Export window. You may export and save protocols to external storage media.

Page 39 -  Closing SonoView

7-93Chapter 7 UtilityStarting a Protocol and Capturing ImagesAfter registering a patient ID, press the Stress Echo button in the Utility menu. The Pr

Page 40

7-94Operation ManualTips!Best Beat SelectionWhen you select Best Beat as the Capture Unit and press the Save button, a screen appears for selecting th

Page 41 - Chapter 7

7-95Chapter 7 Utility Begin Protocol/End ProtocolBegin or end a protocol. NOTE: A protocol cannot be started again once it has ended.  Pause Pr

Page 42

7-96Operation Manual  Icon List [Figure 7.55 Icon List]1 Assign: Click the icon to turn on or o. When this is turned on, the stage selected in Sta

Page 43 - System Settings

7-97Chapter 7 UtilityReviewing an ImageSelect a Stress Echo image on the Protocol Screen, and press Review. The Stress Echo Review screen will appear

Page 44 -  Clip Store Setting

7-98Operation Manual[Figure 7.56 WMS diagram]Left Ventricular Segment Area NumberBasal Anterior Septum 1Basal Anterior 2Basal Lateral 3Basal Posterior

Page 45 -  Scan Mode

7-99Chapter 7 Utility Entering WMS1. Select the segment that you want to enter a WMS into. The Score menu will appear.2. Select a value from the

Page 46 -  Control

6-6Operation Manual Annotating Images TextYou can enter comments or text over an image. This function can be useful when the diagnosis area is diere

Page 47 - Display Settings

7-100Operation Manual Delete WMSDelete all WMS entered in the current stage.  Run/StopAll the Cine images in the Stress Echo image area are played

Page 48

7-101Chapter 7 Utility Strain Image (Optional)The Strain Image function shows heart motion quantitatively using a motion vector, and diagnoses heart

Page 49

7-102Operation ManualQuanticationPress the Soft Menu dial-button [1] Quantication to select Auto EF, Strain menus. NOTE: If the image is not properl

Page 50

7-103Chapter 7 UtilityXDraw contour: This will be displayed on images that do not contain Strain data. Press this button, and then use the control p

Page 51 - Annotate Settings

7-104Operation ManualStrain ReviewPress the Soft Menu dial-button [6] Strain to enter the Strain Result screen. 12384567[Figure 7.62 Strain Result Sc

Page 52

7-105Chapter 7 Utility  Parametric ImageProvides four Parametric Images, which represent the vector direction, velocity, strain, and movement positi

Page 53 - Restart Position

7-106Operation Manual  M ModeGives a color representation of all of the contour components. Place the pointer over the M Mode graph. A white line wil

Page 54

7-107Chapter 7 Utility  Measure ResultReview the measurement results together. Check the check box for the item you have nished reviewing to approv

Page 55 - Peripherals Setup

7-108Operation Manual  User Dened LayoutPosition the on-screen elements in the layout of your choice.  Layout SettingPress the Soft Menu dial-butt

Page 56 -  Foot Switch

7-109Chapter 7 UtilityAuto EFPress the Soft Menu dial-button [1] Quantication button to dene a contour line, and press the 6 Auto EF button. NOTE:

Page 57

6-7Chapter 6 Image Management  Typing TextUse the Alphanumeric Keyboard. Use the Document menu or the Soft Menu dial-button [3] to enter text automa

Page 58 - System Information

7-110Operation Manual  Auto EF Result 1 23[Figure 7.63 Auto EF Result Screen] The Auto EF Result screen consists of the following items: 1 End Diasto

Page 59 - DICOM Setup (Optional)

Chapter 8MaintenanceChapter8 MaintenanceSystem Maintenance ...8-3Installation Requirements ...

Page 61 -  Adding DICOM Service

8-3Chapter 8 Maintenance System MaintenanceNOTE: The user must ensure that the safety inspections are performed every 2 years according to the req

Page 62 -  Storage Server Information

8-4Operation ManualCleaning and disinfections Using inappropriate detergent or disinfectant may damage the product. Please read the following carefull

Page 63 -  Print Server Information

8-5Chapter 8 Maintenance  DisinfectionCAUTION: Use only recommended disinfectants on system surfaces.A disinfectant qualied by the FDA 510(k) pro

Page 64

8-6Operation ManualFuse ReplacementThe power protection fuse protects the product from excess current. If the power monitoring protection circuit dete

Page 65 - Conguration Info

8-7Chapter 8 Maintenance Cleaning The Air FiltersThe air lters minimize the indraft of dust. Clean the air lter to ensure that a clogged lter doe

Page 66 - Start Date

8-8Operation ManualAccuracy CheckThe product’s maintenance status may aect the measurements obtained using the product. The product should be maintai

Page 67

8-9Chapter 8 Maintenance Administration of InformationCAUTION: You may lose information files on user settings or patients, because of shock on th

Page 68 -  PPS Server Information

6-8Operation Manual[Figure 6.3 Text Edit] Change App.Use the Soft Menu dial-button [5]. Change the application by rotating the dial-button. As the a

Page 70 -  Testing DICOM Server

Chapter 9Probes ...9-3Ultrasound transmission Gel ...

Page 72 - Number of Jobs

9-3Chapter 9 Probes ProbesThe probe is a device that sends and receives ultrasound for acquiring image data. It is also called a Transducer or Scan

Page 73 -  DICOM Log

9-4Operation ManualProbe Application Preset TypeCF4-9AbdomenGeneral, Renal, AortaMini DLPPediatricGeneral, Neo headVascularCarotid, Arterial, VenousC4

Page 74

9-5Chapter 9 Probes Probe Application Preset TypeEVN4-9OB1st TrimesterMini DLPGynecologyUterus, AdnexaUrologyProstateER4-9/10EDOB1st Trimester156 Pi

Page 75 - Utilities Setup

9-6Operation ManualProbe Application Preset Type3DC2-6AbdomenGeneral, Renal, AortaMini DLP / 260 PinOBFetal Heart, 1st Trimester, 2nd / 3rd TrimesterG

Page 76 -  Set / Exit Key Setup

9-7Chapter 9 Probes  Function listThe modes and functions that can be used according to the probes and applications in the SONOACE R7 are as follo

Page 77 -  e-Motion Marker

9-8Operation ManualProbe Application HAR PI PPI SCI DMR+ Q Scan ECGL5-12/50EPSmall Parts O X O O O X156 PinVascular O X O O O OMusculoskeletal O X O O

Page 78 - Option Setup

9-9Chapter 9 Probes Probe Application HAR PI PPI SCI DMR+ Q Scan ECGSP3-8Abdomen O O X O O XMini DLPCardiac O O X O O OTCD O O X O O XP2-4AHAbdomen

Page 79 - Auto Calc

6-9Chapter 6 Image ManagementBodyMarker You can place a BodyMarker over an image. This function can be useful when the diagnosis area is dierentiat

Page 80 - Measurement Setup

9-10Operation ManualProbe Application HAR PI PPI SCI DMR+ Q Scan ECG3D4-9ESOB O O X O O O260 PinGynecology O O X O O OUrology O O X O O OC W2.0 Cardia

Page 81 -  Cursor & Method

9-11Chapter 9 Probes Probe Application C/M TDI PD DPDI TDW CW BiopsyC2-5Abdomen X X O O X OMini DLP / 156 PinOBX(Except Fetal Heart)X O O X OGynecol

Page 82 -  Start Point of Measurement

9-12Operation ManualProbe Application C/M TDI PD DPDI TDW CW BiopsyL5-12/50EPSmall Parts X X O O X O156 PinVascular X X O O X OMusculoskeletal X X O O

Page 83

9-13Chapter 9 Probes Probe Application C/M TDI PD DPDI TDW CW BiopsySP3-8Abdomen X X O O X XMini DLPCardiac O O X X O XTCD X X O O X XP2-4AHAbdomen

Page 84 -  Calc. Package(s) Order

9-14Operation ManualProbe Application C/M TDI PD DPDI TDW CW Biopsy3D4-9ESOB X X O O X O260 PinGynecology X X O O X OUrology X X O O X OCW2.0 Cardiac

Page 85 - Link with application

9-15Chapter 9 Probes  Thermal Index (TI) TablesTI values displayed on the screen title bar can change depending on probes and applications. SONOAC

Page 86 -  Custom Menu

9-16Operation ManualProbe Application Preset TI TypeL5-12/50Small Parts Thyroid, Breast, Testicle, Bowel TISMini DLPVascular Carotid, Arterial, Venous

Page 87 - Available Menu List

9-17Chapter 9 Probes Probe Application Preset TI TypeP2-4Abdomen General, Renal, Aorta TISMini DLPCardiac Aortic arch, Adult Echo, Pediatric Echo TI

Page 88

9-18Operation ManualProbe Application Preset TI TypeVN4-8Abdomen General, Renal, Aorta TISMini DLPOB1st Trimester TISFetal Heart, 2nd / 3rd Trimester

Page 89 -  Report

9-19Chapter 9 Probes Ultrasound transmission GelFor proper transmission of the acoustic beam, only use ultrasound transmission gel only approved by

Page 90 -  Print Format

6-10Operation Manual  Entering BodyMarker 1. BodyMarkers will appear on the Soft Menu. X The types of BodyMarker shown in the Soft Menu vary with th

Page 91 -  Data Transfer

9-20Operation ManualSheathsSheaths are recommended for clinical applications of an invasive nature, including intraoperative, transrectal, transvagina

Page 92 -  Caliper

9-21Chapter 9 Probes Probe PrecautionsCAUTION: XDo not apply mechanical shock to the probe.XDo not place the probe cable on the floor where the ca

Page 93

9-22Operation Manual  Electric ShocksThe probe uses electrical energy. If it touches conductive materials, there are risks of electric shocks to the

Page 94 - Obstetrics Measurement Setup

9-23Chapter 9 Probes Cleaning and Disinfecting the ProbeUsing an inappropriate detergent or disinfectant may damage the probe.WARNING: XAlways use

Page 95 -  Percentile Info

9-24Operation ManualNamesDisinfectantsT-Spray IIT-SpraySani-Cloth HBSani-Cloth PlusSani-Cloth ActiveSeptiwipesClieanisept WipesSter-Bac BluTranseptic

Page 96 -  Table Settings

9-25Chapter 9 Probes NamesDisinfectantsCidex OPA 2,3,6)Cidex Plus 2)Metricide 2)Omnicide (28)Omnicide 14NSOmnicide - FG2NucleanWavicide-01 3)Sekusep

Page 97 - Add Reference

9-26Operation ManualNamesDisinfectantsCidex OPA 2,3,6)Cidex Plus 2)Metricide 2)Omnicide (28)Omnicide 14NSOmnicide - FG2NucleanWavicide-01 3)Sekusept E

Page 98

9-27Chapter 9 Probes NamesDisinfectants CleanerGigaseptGigasept AF 3)Gigasept FFHibitanePeraSaveEnzolAlkazymeCidezymeKlenzymeIsopropyl alcohol(70%)I

Page 99

9-28Operation ManualNamesDisinfectants CleanerGigaseptGigasept AF 3)Gigasept FFHibitanePeraSaveEnzolAlkazymeCidezymeKlenzymeIsopropyl alcohol(70%)Isop

Page 100 - View & Modify Reference

9-29Chapter 9 Probes NamesGelNatural ImageAquasonics 100 3)GE Ultrasound Contact GelClear ImageKendallScanWavelengthSonogelType G G G G G G G GActiv

Page 101 -  Trend graph

6-11Chapter 6 Image ManagementIndicator You can place an Indicator over an image. This function can be useful when the diagnosis area is dierentiate

Page 102 - Cardiac Measurement Setup

9-30Operation ManualNamesGelNatural ImageAquasonics 100 3)GE Ultrasound Contact GelClear ImageKendallScanWavelengthSonogelType G G G G G G G GActive I

Page 103 - Vascular Measurement Setup

9-31Chapter 9 Probes Tips!Meanings of the SymbolsThe meanings of the symbols in the table are as follows. (1) Compatible but no EPA Registration

Page 104 - Urology Measurement Setup

9-32Operation ManualThe information on the manufacturers of cleaning agents, disinfectants and ultrasound gel is as follows.Product Manufacturer or Di

Page 105 -  Display

9-33Chapter 9 Probes  CleaningCleaning is an important procedure that is carried out before disinfecting the probe. The probe must be cleaned afte

Page 106 - Fetal Echo Measurement Setup

9-34Operation ManualCAUTION: XUsing a non-recommended disinfectant or not following the recommended disinfection method can damage and/or discolor t

Page 107 - Utility

9-35Chapter 9 Probes BiopsyA biopsy is an examination method that surgically extracts tissue from the patient for examination. The probe and the bi

Page 108 -  Soft Menu

9-36Operation ManualUsing the Biopsy KitWARNING: XVerify the condition of the biopsy needle before use. Do not use a bent biopsy needle.XThe biopsy

Page 109 -  ECG Setup

9-37Chapter 9 Probes 5. Insert the needle into the needle guide. Perform the puncture by sliding the needle through the groove in the guide until t

Page 110 - Gain / Auto Gain

9-38Operation ManualCleaning and Disinfecting Biopsy KitWash and disinfect the biopsy kit to reduce pathogens to the level of 10-6. Some components of

Page 111 - Chapter 7 Utility

9-39Chapter 9 Probes  DisinfectionCAUTION: Plastic biopsy kits can only be disinfected using a chemically compatible cold-disinfectant. Disinfecti

Page 113 - Histogram

6-12Operation Manual  Deleting Indicator Press the Clear button on the control panel and all Indicators displayed on the screen will be deleted. Pres

Page 114 - Operation Manual

9-40Operation ManualAssembling the Biopsy Kit  C2-5 Biopsy Guide (Plastic)  SpecicationsModel NameBP-KIT-045ComponentsBiopsy Adapter 1 EA (Reusab

Page 115 - Post Curve

9-41Chapter 9 Probes 3. Secure the Needle Guide Clip.4. Insert the needle into the needle guide for use.

Page 116 -  Color Map

9-42Operation Manual  C2-5 Biopsy Guide (Metal)  SpecicationsModel NameBP-KIT-042ComponentsBiopsy Adapter 1 EA Needle Guide 24 EAMaterialStainle

Page 117 -  D Post

9-43Chapter 9 Probes 3. Secure the Needle Guide Clip.4. Insert the needle into the needle guide for use.

Page 118 -  Monitor Calibration

9-44Operation Manual  C2-8 Biopsy Guide  SpecicationsModel NameBP-KIT-035ComponentsBiopsy Adapter 1 EA (Reusable)Needle Guide 24 EA (Disposable)

Page 119 -  Gamma

9-45Chapter 9 Probes 3. Mount the needle guide for the desired gauge onto the adapter.4. Insert the needle into the needle guide for use.

Page 120 - Measure Setup

9-46Operation Manual  L3-8 Biopsy Guide  SpecicationsModel Name.BP-KIT-036ComponentsBiopsy Adapter 1 EA (Reusable)Needle Guide 24 EA (Disposable

Page 121 - Storage Manager

9-47Chapter 9 Probes 3. Mount the needle guide for the desired gauge onto the adapter.4. Insert the needle into the needle guide for use.

Page 122 - Eject / Remove

9-48Operation Manual  L5-12/50 Biopsy Guide (Plastic) SpecicationsModel NameBP-KIT-047ComponentBiopsy Adapter 1 EA (Reusable)Needle Guide 24 EA

Page 123 - BodyMarker Edit

9-49Chapter 9 Probes 3. Secure the Needle Guide Clip.4. Insert the needle into the needle guide for use.Tips!Multi AngleUse the Angle-adjusting Me

Page 124 - Resetting the BodyMarker list

6-13Chapter 6 Image Managemente-Motion Marker (Optional)e-Motion Marker displays the probe's position and angle on the screen at the location be

Page 125 - Button Setup

9-50Operation Manual  HL5-12ED Biopsy Guide SpecicationsModel Name.BP-KIT-022 ComponentsBiopsy Adapter 1 EA (Reusable)Needle Guide 24 EA (Dispos

Page 126 - Menu Edit

9-51Chapter 9 Probes 3. Mount the needle guide for the desired gauge onto the adapter.4. Lock the needle guide as in the following (a), (b). and (

Page 127

9-52Operation Manual  L5-12/50EP Biopsy Guide  SpecicationModel NameBP-KIT-020ComponentsBiopsy Adapter 1 EA (Reusable)Needle Guide 24 EA (Dispos

Page 128 - Stress Echo (Optional)

9-53Chapter 9 Probes 2. Mount the needle guide for the desired gauge onto the adapter.3. Lock the needle guide. 4. Insert the needle into the nee

Page 129 - Setting Up a Protocol

9-54Operation Manual  LN5-12 Biopsy Guide  SpecicationModel nameBP-KIT-040ComponentBiopsy Adapter 1 EA (Reusable)Needle Guide 24 EA (Disposable)

Page 130 - Protocol

9-55Chapter 9 Probes 3. Secure the Needle Guide Clip.4. Insert the needle into the needle guide for use.

Page 131

9-56Operation Manual  EVN4-9, ER4-9, ER4-9/10ED, EV4-9/10ED Biopsy Guide SpecicationsModel NameBP-KIT-024ComponentsBiopsy Adapter 1 EASheath 2 E

Page 132

9-57Chapter 9 Probes 3. Place the sheath again and tie with the rubber band again. Tie the lower part than 1. 4. Insert the needle into the needle

Page 133 - Capturing Stress Echo Images

9-58Operation Manual  3DC2-6 Biopsy Guide SpecicationsModel NameBP-KIT-026ComponentsBiopsy Adapter 1 EA (Reusable)Needle Guide 24 EA (Disposable

Page 134 -  Protocol Screen

9-59Chapter 9 Probes 3. Mount the needle guide on the adapter.4. Install the needle into the needle guide for use.

Page 135

6-14Operation Manual  e-Motion Marker SettingGo to Utility > Setup > Utility > e-Motion Marker and press the Control button to open the setu

Page 136 -  Icon List

9-60Operation Manual  VN4-8 Biopsy Guide (plastic) SpecicationsModel NameBP-KIT-049ComponentBiopsy Adapter 1 EA (Reusable)Needle Guide 24 EA (Di

Page 137 - Reviewing an Image

9-61Chapter 9 Probes 3. Secure the Needle Guide Clip.4. Insert the needle into the needle guide for use.Tips!Multi AngleUse the Angle-adjusting Me

Page 138

9-62Operation Manual  3D4-8, 3D4-8ET Biopsy Guide  SpecicationsModel NameBP-KIT-033ComponentsBiopsy Adapter 1 EA (Reusable)Needle Guide 24 EA (

Page 139 - Entering WMS

9-63Chapter 9 Probes 3. Mount the needle guide.4. Insert the needle into the needle guide for use.

Page 140

9-64Operation Manual  3D4-9 Biopsy guide SpecicationsModel nameBP-KIT-027ComponentBiopsy Adapter 1EACleaning Brush 5EAMaterialStainlessAvailable

Page 141 - Strain Image (Optional)

6-15Chapter 6 Image Management Button SetupCongure the User Key for e-Motion Marker.XButton 1: This is the Initial key, which initiates the e-Mot

Page 142 - Quantication

6-16Operation Manual Saving, Playing and Transferring ImagesSaving Images  Saving Still ImagesPress the Save button on the control panel. Images a

Page 143 - Modifying a Contour

6-17Chapter 6 Image Management Method X Manual: Specify the start and end of an image by using the Store Clip button. That is, the image that is d

Page 144 - Strain Review

6-18Operation ManualPlaying ImagesThe saved images can be played in SonoView or in a diagnosis mode.  Playing Images in SonoViewNOTE: Please refer t

Page 145 -  Segmental Graph

6-19Chapter 6 Image ManagementTransferring Images SONOACE R7 allows you to transfer images to the PACS system supporting DICOM. You can transfer all

Page 146 -  Analysis Type

6-20Operation Manual Printing and Recording Images Printing ImagesPress the Print button on the control panel. Images are printed via an echo printer

Page 147 -  User Preset

6-21Chapter 6 Image Management SonoViewSonoView is an integrated image management program. It provides a range of functions including: saving/ling o

Page 148 -  Exiting Strain

Operation ManualVersion 3.01.00EnglishDiagnostic Ultrasound SystemMI68-02773A

Page 149 - Calculating Auto EF Data

6-22Operation Manual  Image Scroll The image of the previous or next page is displayed on the screen. Use the Soft Menu dial-button [2] Image Scroll

Page 150 -  Auto EF Result

6-23Chapter 6 Image ManagementCompare ModePress Compare in the upper left corner of the screen. The button will appear in yellow.In Compare Mode, you

Page 151 - Maintenance

6-24Operation Manual  Selecting ImagesUse the Trackball and the Set button to select an image from the thumbnail list. The selected image is highligh

Page 152

6-25Chapter 6 Image Management  Closing Exam ReviewPress Close on the screen. To close all exams in the list, press the Close All button.  LayoutsY

Page 153 - System Maintenance

6-26Operation Manual  Reviewing 3D ImagesIf the saved image is 3D, the 3D indication ( ) appears at the bottom of the image. Press 3D and the 3D View

Page 154 - Cleaning and disinfections

6-27Chapter 6 Image Management  Measuring Circumference and AreaPress the Ellipse Calliper ( ) icon on the screen. You can measure the circumference

Page 155 -  Disinfection

6-28Operation Manual[Figure 6.13 Image Print]  Transferring Images via DICOMYou can send the selected images via DICOM. The icon is enabled only when

Page 156 - Fuse Replacement

6-29Chapter 6 Image Management  Printing via DICOMYou can print the selected images via DICOM. The icon is enabled only when an image(s) is selected

Page 157 - Cleaning The Air Filters

6-30Operation Manual[Figure 6.16 Image Export]  Sending in E-mail1. Press the Send E-mail ( ) icon on the screen. The E-mail window will appear.2. S

Page 158 - Accuracy Check

6-31Chapter 6 Image Management[Figure 6.17 E-mail]  Deleting ImagesPress the Delete Image ( ) icon on the screen. Press OK to delete the selected im

Page 161 - Chapter 9

Chapter 7System Settings ...7-3General System Settings ...

Page 162

Chapter 7Utility ... 7-67ECG ...

Page 163 -  Probe List

7-3Chapter 7 Utility System SettingsThis mode is used for system settings. It does not aect image output. The setup may be modied depending on spe

Page 164

7-4Operation ManualNOTE: XYou cannot change the date and time when a patient ID is registered. To change the date and time, you should finish the cur

Page 165 - Chapter 9 Probes

7-5Chapter 7 UtilityX ECG Beat: Specify the heart beat as 1 – 8 beats.X Time: Specify it as 1 – 50 secondsX Manual: Save images automatically for

Page 166

7-6Operation Manual End Exam ActionXEnd Exam Only: Pressing the End Exam button on the control panel exits Exam Mode and switches to the B Mode Sca

Page 167 - Function list

7-7Chapter 7 UtilityDisplay SettingsSelect the Display tab in the Setup screen. Specify display-related options. [Figure 7.3 Setup -Display]

Page 168

7-8Operation Manual  Display Option You can specify more than one item. Use the Trackball and the Set button to select a desired item and check or

Page 169

7-9Chapter 7 Utility 2D/C LiveSpecify how Color Doppler Mode is displayed in 2D/C Live Mode. In the 2D menu, if Horizontal Dual is On, it can be pl

Page 170

5Table of ContentsTable of Contents – Volume 2Chapter6 Image ManagementReviewing Images (Cine / Loop) ...

Page 171

7-10Operation Manual PreviewPreview window displays the font you selected.DefaultGo back to the default setting like below.Document Font Measure R

Page 172

7-11Chapter 7 UtilityAnnotate SettingsSelect the Annotate tab in the Setup screen. Congure the information related to image annotation.  BodyMarker

Page 173

7-12Operation ManualTips!Editing Abbreviation ListTo enable the abbreviation list stored in the system, press the Autotext Edit button. The system wil

Page 174

7-13Chapter 7 Utility Quick TextIf the checkbox is selected, the Quick Text function is enabled. Quick Text switches the system to the Text Input M

Page 175 -  Thermal Index (TI) Tables

7-14Operation Manual[Figure 7.6 Annotate]

Page 176

7-15Chapter 7 UtilityPeripherals SetupSelect the Peripherals tab in the Setup screen. You can congure keys, buttons and the peripheral devices conne

Page 177

7-16Operation Manual Local Printing AreaSet the area that will be printed.X Full Screen (1280 * 1024): Print the full monitor screen. X Video Out

Page 178

7-17Chapter 7 Utility[Figure 7.7 Setup - Peripherals]

Page 179 - Ultrasound transmission Gel

7-18Operation ManualSystem Information Select the Info tab in the Setup screen. Information about the system S/W version will be displayed. Press Deta

Page 180 -  Installing the Sheath

7-19Chapter 7 UtilityDICOM Setup (Optional)Select the DICOM tab in the Setup screen. You can congure the DICOM (Digital Imaging and Communication in

Page 181 - Probe Precautions

6Operation ManualMeasurement Setup ...7-4

Page 182 -  Electric Shocks

7-20Operation Manual AE TitleEnter the name of the DICOM AE (Application Entity). The title is used to identify devices that use DICOM in a network.

Page 183

7-21Chapter 7 UtilityTips!Display CompensationAdjusting the post curve settings for images allows other monitors to display them as close as possible

Page 184

7-22Operation Manual Connect TimeoutSpecify how long the system will wait until it receives a response from the DICOM server. You can specify it in

Page 185

7-23Chapter 7 UtilityX Window Center: Enter a value for the DICOM Tag (0028, 1050) setting. The setting value indicates the brightness of an image t

Page 186

7-24Operation Manual FormatSpecify the paper layout. Select from 1ⅹ1, 1ⅹ2, 2ⅹ2, 2ⅹ3, 3ⅹ3, 3ⅹ4, 3ⅹ5, 4ⅹ4, 4ⅹ5 and 4ⅹ6. OrientationSpecify the paper

Page 187

7-25Chapter 7 Utility DestinationSpecify the paper pathway. Select Magazine or Processor.Smoothing TypeThis option is available only when Magnic

Page 188

7-26Operation Manual  Worklist Server InformationSelect WORKLIST under Services. Congure the Modality Worklist Service using DICOM. Update Method

Page 189

7-27Chapter 7 Utility[Figure 7.13 DICOM Configuration -Worklist]

Page 190

7-28Operation Manual  PPS Server InformationSelect PPS (Performed Procedure Step) under Services. Congure the Modality Performed Procedure Step Serv

Page 191

7-29Chapter 7 Utility  SC Server InformationSelect SC (Storage Commitment) under Services. Congure the Storage Commitment Service using DICOM. The

Page 192

7Table of ContentsChapter8 MaintenanceSystem Maintenance ...

Page 193 -  Cleaning

7-30Operation Manual  Storage SR Server InformationSelect Storage SR (Storage Structured Report) under Services. Congure the Report Storage Service

Page 194

7-31Chapter 7 Utility  Managing DICOMClick the Queue on the screen to switch to the DICOM Job Status screen. You can review the current job status u

Page 195 - Biopsy KIT Components

7-32Operation Manual6. Date/Time: Display the date and time when the job was created.7. Status: Display the current status of the job.Status Descrip

Page 196 - Using the Biopsy Kit

7-33Chapter 7 Utility Retry AllRetry all jobs for which the status is Fail.DeleteDelete the selected job. This button is enabled only when the st

Page 197 -  Needle Guide Alignment

7-34Operation Manual[Figure 7.18 DICOM Log]

Page 198 - Disinfection

7-35Chapter 7 UtilityUtilities SetupSelect the Utility tab in the Setup screen. You can congure settings for e-mail, text and network status.[Figure

Page 199

7-36Operation Manual  Buzzer Control Specify whether a buzzer sound will be generated when a button or dial-button is pressed. Buzzer SoundIf Buzze

Page 200 - Assembling the Biopsy Kit

7-37Chapter 7 Utility  Network CongurationPress the Network Conguration button, and the TCP/IP Properties… window will open to let you congure th

Page 201

7-38Operation ManualOption SetupSelect the Option tab in the Setup screen. You can enable or disable optional software and hardware.  OptionsThe lis

Page 202 -  C2-5 Biopsy Guide (Metal)

7-39Chapter 7 UtilityAuto CalcAuto Calc is a Spectral Doppler mode feature that automatically performs specic calculations based on measured values.

Page 204 -  C2-8 Biopsy Guide

7-40Operation Manual Measurement SetupSpecify various setup options for measurement. The setup may be modied depending on specic needs or preferenc

Page 205

7-41Chapter 7 Utility[Figure 7.22 General - General]  Cursor & Method Line Marker TypeSpecify the shape of the caliper cursor displayed on the

Page 206 -  L3-8 Biopsy Guide

7-42Operation Manual  Line TypeFrom the following three options, select the line pattern to use when measuring a distance.Dotted LineDisplay with d

Page 207

7-43Chapter 7 Utility Display the M-Mode GuidelineSpecify whether to display cross lines while items are measured in the frozen state in M Mode. Th

Page 208 - Installation

7-44Operation Manual  PackagesSelect the Packages tab under the General tab. Specify the measurement packages that will be used and the order in whic

Page 209

7-45Chapter 7 Utility Adding a Calculation Package to UseSelect a calculation package you want to use from the right list and then click on the s

Page 210 -  HL5-12ED Biopsy Guide

7-46Operation ManualProbe Application Measure ApplicationThyroidThyroidBowelBreast BreastTesticle TesticleAortic archCardiacAdult EchoPediatric EchoPr

Page 211

7-47Chapter 7 UtilityThe following describes custom menu screen.Package ListThe calculation packages set in the Package screen appear in this list.

Page 212 -  L5-12/50EP Biopsy Guide

7-48Operation ManualThe following describes how to set a custom menu.1. In Package List, select an application.2. In Tab List, select a tab menu. Th

Page 213

7-49Chapter 7 Utility  ReportSelect the sub tab Report under the General tab. You can specify options for measurement reports and printing.[Figure 7

Page 214 -  LN5-12 Biopsy Guide

Chapter 6Reviewing Images (Cine / Loop) ...6-3Annotating Images ...6-6Text ...

Page 215

7-50Operation Manual  Report HeaderSpecify header options for reports. You can specify more than one item and the specied items appear in all measur

Page 216

7-51Chapter 7 Utility  OB / FH Header LayoutSpecify item(s) to display under the header of obstetrics or fetal echo measurement reports. You can spe

Page 217

7-52Operation Manual[Figure 7.26 General - Data Transfer]  CaliperSelect the Caliper sub-tab under the General tab. Specify whether additional inform

Page 218 -  3DC2-6 Biopsy Guide

7-53Chapter 7 Utility[Figure 7.27 General - Caliper]  Display Num of 2D Mode Result BoxSpecify the number of lines for 2D Mode results that are di

Page 219

7-54Operation ManualObstetrics Measurement SetupSelect the OB tab in the Measure Setup screen. You can specify obstetrics options.  General Settings

Page 220

7-55Chapter 7 Utility  Percentile Info. Show Percentile Info.Specify whether to use Percentile Info. by using this checkbox. Percentile Criteria

Page 221

7-56Operation Manual  Fetal Weight UnitSpecify the unit for fetal weight measurement. You can select up to 2 units from grams [g], ounces [oz] and po

Page 222

7-57Chapter 7 Utility EFW Sequential MeasurementNOTE: These settings are needed when User Key 1 is set to EFW Measure in Utility > Setup > Pe

Page 223

7-58Operation Manual1. Press . The User Reference window will appear.[Figure 7.30 User Reference]2. Enter a name and description for a new referenc

Page 224 -  3D4-9 Biopsy guide

7-59Chapter 7 UtilityTips!Add Reference Table Selection A preset for the selected measurement reference is displayed. The name and type of the refere

Comments to this Manuals

No comments