Operation Manual
Approved Accessories for Electromagnetic Compliance
Accessories used with this product may eect its emissions.
CAUTION: When connecting other customer-supplied accessories to the system, it is the users
responsibility to ensure the electromagnetic compatibility of the system. Use only CISPR 11 or CISPR
22, CLASS B compliant devices.
WARNING: The use of cables, transducers, and accessories other than those specified may result in
increased emission or decreased Immunity of the Ultrasound System
Immunity test IEC 60601 Test level Compliance level
Electromagnetic
environment -guidance
Electrotatic
discharge (ESD)
IEC 61000-4-2
±6KV Contact
±8KV air
±6KV Contact
±8KV air
Floors should be wood,
concrete or ceramic tile.
If oors are covered with
synthetic material, the relat-ive
humidity should be at least
30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2KV
for power supply lines
±1KV
for input/output lines
±2KV
for power supply lines
±1KV
for input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1KV dierential mode
±2KV common mode
±1KV dierential mode
±2KV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% Uт for 0.5 cycles
(>95% dip in Uт)
40% Uт for 5 cycles
(60% dip in Uт )
70% Uт for 25 cycles
(30% dip in Uт)
<5% Uт for 5 s
(<95% dip in Uт )
<5% Uт for 0.5 cycles
(>95% dip in Uт)
40% Uт for 5 cycles
(60% dip in Uт )
70% Uт for 25 cycles
(30% dip in Uт)
<5% Uт for 5 s
(<95% dip in Uт )
Mains power quality should be
that of a typical commercial
or hospital environment. If the
user of this pro;duct requires
continued operation during
power mains interrup-tions,
it is recommended that this
product be powered from an
uninterruptible power supply
or a battery.
Power frequency
(50/60Hz)
magnetic eld
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic
elds should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE Uт is the a.c. mains voltage prior to application of the test level.