Volume 2
7-64Operation ManualConfigure the calculation menu as follows:1. Select an application under Select Calc Package.2. Select a menu tab under Check Ca
7-65Chapter 7 Utilities [Figure 7.36 General – Calc Menu]
7-66Operation ManualReportSelect the Report tab under the General tab. Here you can set items related to the measurement report and printing.[Figure 7
7-67Chapter 7 Utilities Measurement Result Measurement Result TypeUsed to specify the method of calculating measurement values in the reports.X
7-68Operation ManualData TransferSelect the sub-tab Data Transfer under the General tab. Here you can set the data transfer method, etc.[Figure 7.38 G
7-69Chapter 7 Utilities CaliperSelect the Caliper sub-tab under the General tab. Specify whether additional information will be shown along with the
7-70Operation ManualOBSelect the OB tab from the Measure Setup screen and then configure OB measurement-related settings. GeneralSelect the sub-tab Ge
7-71Chapter 7 Utilities Rank Information MethodSpecify how the growth range information will be displayed. The growth range information can be us
7-72Operation Manual Number of Measure Results Displayed Number of Results DisplayedSpecify the number of lines for OB measurement results that are
7-73Chapter 7 Utilities TablesSelect the sub-tab Tables under the OB tab. You can specify references such as reference tables and equations that wil
6-3Chapter 6 Image Management Cine/LoopImages are automatically saved to the system during scanning. Saved images can then be used to diagnose the
7-74Operation Manual Add a ReferenceNOTE: Observe the following directions when adding a table reference. If these conditions are not met, a warning
7-75Chapter 7 Utilities Tips!Add Reference TableClicking the Question Mark button shows the sources of the references. Clicking the question mark bu
7-76Operation ManualTips!Adding Reference EquationIf a reference appears in an equation, the following should be entered:EquationEnter a reference equ
7-77Chapter 7 Utilities Calc & GraphSelect the sub-tab Calc & Graph under the OB tab. You can specify settings for calculation and graphs.
7-78Operation ManualCardiacSelect the Cardiac tab from the Measure Setup screen and then configure cardiac measurement-related settings. Cursor &
7-79Chapter 7 Utilities Type of Derived Calc Results DisplayedUsed to specify the measurement result display method. Select Brief for a brief displ
7-80Operation ManualVascularSelect the Vascular tab on the Measure Setup screen. You can specify settings for vascular measurement. [Figure 7.47 Vascu
7-81Chapter 7 Utilities UrologySelect the Urology tab on the Measure Setup screen. You can specify settings for urology measurement. [Figure 7.48 Ur
7-82Operation Manual Predicted PSA correction factorSpecify the predicted Prostate Specific Antigen (PSA) correction factor for the measurement of WG
7-83Chapter 7 Utilities Fetal HeartSelect the Fetal Heart tab on the Measurement Setup screen. Here you can set items related to fetal heart measure
6-4Operation Manual Starting and Finishing Image ReviewDuring scanning, press the Freeze button on the control panel. This stops the scan and switches
7-84Operation Manual FH Measurement Result Measurement Result TypeSpecify the measurement method. Select Avg, Last, Max, or Min value. Number of Mea
7-85Chapter 7 Utilities Storage ManagerSelect Storage Manager from the utility menu. All disk drives mounted in the system will be shown. The driv
7-86Operation Manual Eject / RemoveDisconnects the selected drive. NOTE: Before unplugging a USB Flash memory drive, make sure to disconnect it by u
7-87Chapter 7 Utilities Menu EditSelect Menu Edit from the utility menu. The Menu Edit screen will appear. From this screen, you can configure the
7-88Operation ManualMode Item2D Mode2D Image SizeAngle (10 Steps)Chroma MapDMR+ IndexDynamic RangeEdge EnhanceFocusFrame AvgFrame RateFSIGray MapLine
7-89Chapter 7 Utilities Mode ItemCW ModeAngle (10 Steps)AutoCalc DirectionBaseLine Chroma MapDisplay FormatDynamic RangeLoop SizePower ScaleSpectrum
7-90Operation Manual[Figure 7.52 User Presets] − Load: Applies the selected User Preset to the system. − Save: Saves the current settings to the sele
Operating Environment ...8-3System Maintenance ...8-4Cleaning and disinfecting
8-3Chapter 8 Maintenance Operating EnvironmentWhen installing this product, please pay attention to the following:CAUTION: XPlacing the system ne
6-5Chapter 6 Image Management Trim First You can specify the rst frame of the range that you wish to save as part of the Cine or Loop. Rotate th
8-4Operation Manual System MaintenanceNOTE: The expected service life of this product is 7 years.Cleaning and disinfectingUsing an inappropriate clea
8-5Chapter 8 Maintenance DisinfectingCAUTION: Use only the recommended disinfectants on system surfaces.A disinfectant qualied by the FDA 510(k) p
8-6Operation ManualAccuracy CheckNOTE: The user must ensure that safety inspections are performed every 2 years according to the requirements of safet
8-7Chapter 8 Maintenance Battery Pack ManagementThe battery pack is a consumable, and will lose performance over time. If the battery life becomes
8-8Operation ManualRecharging the Battery PackConnecting the AC adapter automatically begins to recharge the battery pack. The battery pack will be re
8-9Chapter 8 Maintenance Storing the Battery PackIf you are not planning on using your MySono U6 unit, remove the battery pack from the unit and sto
8-10Operation Manual Data ManagementCAUTION: You may lose information files on user settings or patients, because of physical shocks to the product o
Probes ...9-3Ultrasound Transmission Gel ...9-
9-3Chapter 9 Probes ProbesThe probe is a device that sends and receives ultrasound for acquiring image data. It is also called a Transducer or Scan
6-6Operation Manual AnnotationText Allows the user to place text on an image. This function can be useful when the diagnosis area is dierentiated or
9-4Operation ManualProbes Application PresetLN5-12Musculoskeletal General Small Parts General, BreastVascular GeneralP2-4Abdomen GeneralCardiac Gene
9-5Chapter 9 Probes Function listAvailable functions based on the type of probe and application being used in MySono U6 are as follows: Probes Ap
9-6Operation ManualProbes Application HAR PI SCI DMR+ Q-SCAN ECG BiopsyEVN4-9Gynecology X X X O O X OOBX X X O O X OUrologyX X X O O X O3DC2-6AbdomenO
9-7Chapter 9 Probes Probes Application CM TDI PD DPDI TDW CW 3D/4DC2-5AbdomenX X O O X X XGynecologyX X O O X X XOBX (Except Fetal Heart)X OO (Excep
9-8Operation ManualProbes Application CM TDI PD DPDI TDW CW 3D/4DEVN4-9Gynecology X X O O X X XOBX X O O X X XUrologyX X O O X X X3DC2-6AbdomenX X O O
9-9Chapter 9 Probes Thermal Index (TI) TablesTI values displayed on the screen title bar can change depending on probes and applications. The follo
9-10Operation ManualProbes Application PresetsThermal Index (TI) TablesP3-7ACAbdomen General TIsCardiac General TIsTCD General TIsSP3-8Abdomen General
9-11Chapter 9 Probes Ultrasound Transmission GelUsing an inappropriate ultrasound gel may damage the probe. For proper transmission of the acoustic
9-12Operation ManualSheath UseSheaths are recommended for clinical applications of an invasive nature, including intraoperative, transrectal, transvag
9-13Chapter 9 Probes Probe Safety PrecautionsCAUTION: XDo not apply mechanical shock to the probe.XDo not place the probe cable on the floor where
6-7Chapter 6 Image Management Entering and Deleting Text Typing TextUse the keyboard. You can move the cursor by using the Trackball or the arrow
9-14Operation Manual Electric ShocksThe probe uses electrical energy. If it touches conductive materials, there are risks of electric shocks to the pa
9-15Chapter 9 Probes Cleaning and Disinfecting the ProbeUsing an inappropriate cleaning or sterilizing agent may damage the product.WARNING: XAlway
9-16Operation ManualNamesDisinfectantsT-Spray IIT-SpraySani-Cloth HBSani-Cloth PlusSani-Cloth ActiveSeptiwipesCleanisept WipesSter-Bac BluTranseptic S
9-17Chapter 9 Probes NamesDisinfectantsCidex OPA 2,3)Cidex Plus 2)Metricide 2)Omnicide (28)Omnicide 14NSOmnicide - FG2NucleanWavicide-01 3)Sekuse
9-18Operation ManualNamesDisinfectants CleanerVirkonSporoxSporox IIGigaseptGigasept AF 3)Gigasept FFHibitanePeraSafeEnzolAlkazymeCidezymeKlenzymeIsop
9-19Chapter 9 Probes NamesCleaner GelEthanol 75%MetrizymeMcKessonNatural ImageAquasonics 100 3)GE Ultrasound Contact GelClear ImageKendallScanWavele
9-20Operation ManualTips!Symbols Legend(1)Compatible but no EPA Registration(2)FDA 510(k) qualified(3)Has CE mark(4)Discontinued(5)Under DevelopmentSS
9-21Chapter 9 Probes The following is information about the manufacturers (or Distributors) of detergents, disinfectants, and ultrasound gels.Produc
9-22Operation Manual CleaningCleaning is an important procedure that is carried out before disinfecting the probe. The probe must be cleaned after eac
9-23Chapter 9 Probes DisinfectionA 10-6 reduction in pathogens should be reached by following the disinfection procedures in this Manual and using
6-8Operation Manual EditPressing the soft menu’s dial-button [4] brings up the Text Edit screen. Type new text into text elds 1 to 10, or edit exis
9-24Operation Manual BiopsyA biopsy is an examination method that surgically extracts tissue from the patient for examination. The probe and the biops
9-25Chapter 9 Probes Using the Biopsy KitWARNING: XVerify the condition of the biopsy needle before use. Do not use a bent biopsy needle.XThe biop
9-26Operation Manual5. Install the needle into the needle guide for use. Insert the needle until it reaches the examination site. To keep the needle
9-27Chapter 9 Probes Cleaning and Disinfecting the Biopsy KitWash and disinfect the biopsy kit to reduce pathogens to the level of 10-6. Some compon
9-28Operation Manual Cleaning and Disinfecting a plastic biopsy kit Cleaning 1. After use, remove the biopsy kit from the probe.2. Disassemble the
9-29Chapter 9 Probes Assembling the Biopsy Kit C2-5 Biopsy Guide (Plastic) Biopsy SpecicationsModel nameBP-KIT-041 ComponentBiopsy BracketNeedle
9-30Operation Manual3. Mount the needle guide for the desired gauge onto the adaptor.4. Install the needle into the needle guide.
9-31Chapter 9 Probes C2-5 Biopsy Guide (Metal) Biopsy SpecicationsModel nameBP-KIT-042 ComponentBiopsy BracketNeedle Guide(16G)Needle Guide(18G
9-32Operation Manual3. Mount the needle guide for the desired gauge onto the adaptor.4. Install the needle into the needle guide.
9-33Chapter 9 Probes C2-8 Biopsy Guide Biopsy SpecicationsModel nameBP-KIT-035ComponentBiopsy BracketNeedle Guide1 EA24 EAMaterialAcetal Copoly
6-9Chapter 6 Image Management Autotext This function allows you to enter text automatically by using an abbreviation. In this way, you can enter te
9-34Operation Manual3. Mount the needle guide for the desired gauge onto the adaptor.4. Install the needle into the needle guide.
9-35Chapter 9 Probes LN5-12 Biopsy Guide Biopsy SpecicationsModel nameBP-KIT-040ComponentBiopsy Adapter 1 EAMaterialAcetal CopolymerAvailable G
9-36Operation Manual3. Mount the needle guide for the desired gauge onto the adaptor. 4. Install the needle into the needle guide.
9-37Chapter 9 Probes EVN4-9 Biopsy Guide Biopsy SpecicationsModel nameBP-KIT-024ComponentBiopsy AdapterSheath1 EA2 EAMaterialStainlessAvailable
9-38Operation Manual 3DC2-6 Biopsy Guide Biopsy SpecicationsModel nameBP-KIT-026ComponentBiopsy AdapterNeedle Guide1 EA24 EAMaterialAcetal Copolym
9-39Chapter 9 Probes 3. Mount the needle guide on the adapter.4. Install the needle into the needle guide for use.
9-40Operation Manual 3D4-9 Biopsy Guide Biopsy SpecicationsModel nameBP-KIT-027ComponentAdapterCleaning BrushInstruction1 EA5 EA1 EAMaterialStainl
6-10Operation ManualBodyMarkerPlaces BodyMarkers on top of images. This function can be useful when the diagnosis area is dierentiated or displayed.[
6-11Chapter 6 Image Management Inserting BodyMarkers 1. Pressing the BodyMarker key displays the BodyMarker list in the thumbnail area. XThe Body
6-12Operation ManualIndicator You can place an Indicator over an image. This function can be useful when the diagnosis area is dierentiated or displa
6-13Chapter 6 Image Management Deleting IndicatorsWhen you press the Clear button on the control panel, all indicators entered on the screen are cl
6-14Operation Manual Saving and Playing ImagesSaving ImagesWARNING: You must always enter the patient ID, because all images are saved according to p
6-15Chapter 6 Image Management Playing ImagesThe saved images can be played in SONOVIEW or in a diagnosis mode. Viewing Images in SONOVIEWNOTE: Ref
6-16Operation Manual Transferring and Printing ImagesTransferring ImagesThis product allows you to transfer images to PACS systems that support DICOM
6-17Chapter 6 Image Management Printing ImagesPress the Print 1 or Print 2 button. Images are printed via an echo printer. For information on congu
6-18Operation Manual SONOVIEWSONOVIEW is MySono U6’s integrated image management program. It oers image saving, referencing, and deleting functions,
6-19Chapter 6 Image Management Image Scroll The image of the previous or next page is displayed on the screen. Use the soft menu dial-button [2] Im
6-20Operation ManualCompare ModeClick Compare in the upper left corner of the screen. The button will appear in yellow.Compare mode is used to compare
6-21Chapter 6 Image Management Select an imageUse the trackball and the Set button to select an image from the thumbnail list. The selected image i
Diagnostic Ultrasound SystemMI68-02776AOperation ManualVersion 1.01.00English
6-22Operation Manual Closing Exam ReviewPress Close on the screen. To close all exams in the list, click the Close All button. LayoutsYou can adjust t
6-23Chapter 6 Image Management Post ProcessingClick the button located at the lower left corner of an image, and the post-processing menu will ap
6-24Operation ManualNOTE: CINE images can only be played in the 2 x 2 or less layout. Searching Exam 1. Click the Exam Search ( ) icon on the screen.
6-25Chapter 6 Image Management Typing TextClick the Text Annotation icon on the screen. You can enter text into an image.1. Place the cursor ove
6-26Operation Manual Transferring Images via DICOMThis product allows you to transfer selected images via DICOM. The icon is enabled only when an imag
6-27Chapter 6 Image Management Transferring ImagesNOTE: Image transfer will not be performed if there are spaces in the file or directory name.1. P
6-28Operation Manual[Figure 6.14 E-mail] Deleting ImagesClick the Delete Image icon on the screen. Press OK to delete the selected image. Please not
Utilities ...7-3ECG ...
Chapter 7AutoCalc ...7-57About ...
7-3Chapter 7 Utilities Utilities Press the Utility key on the keyboard. The Utility Menu and its associated soft menus appear on the screen. With t
7-4Operation Manual Usersets Displays the supported usersets for the current application. Rotate dial-button [3] to select a userset. Pre. LoadPre
7-5Chapter 7 Utilities ECGSelect ECG from the Utility menu. The image of cardiac pulsation will be displayed; this will be displayed in the menu on
7-6Operation Manual GainAdjust the brightness of an image. Rotate the dial-button clockwise to increase the gain. Rotate the Soft Menu dial-button [
7-7Chapter 7 Utilities BiopsySelect Biopsy from the utility menu. NOTE: The biopsy option is not available with the phased array probe. Editing the
7-8Operation Manual Starting and Finishing a BiopsyNOTE: Make sure to adjust the biopsy guideline before using the biopsy feature.1. Press the soft m
7-9Chapter 7 Utilities Histogram A histogram is a type of graph representing the distribution of echoes. 1. Select Histogram from the utility menu
7-10Operation Manual Histogram Settings You can specify the position of the histogram, and the type of histogram that you want to use. Move Hist. Pr
7-11Chapter 7 Utilities Post Curve Select Post Curve from the utility menu. Here you can set various post maps and gamma values. [Figure 7.5 Post
7-12Operation Manual EditUsed to adjust the user type RGB curve. NOTE: Activated only when Curve is set to User1 through 3.Selecting this option chang
7-13Chapter 7 Utilities CurveUsed to specify the curve type. Use the Ext. Menu dial-button to set the curve to Type1 through 5, or User1 through 3.
5Table of ContentsTable of Contents – Volume 2Chapter 6 Image ManagementCine/Loop ...
7-14Operation ManualGammaSelect Gamma from the post curve menu. Use the Ext. Menu dial-button to adjust the brightness and contrast levels. Select O,
7-15Chapter 7 Utilities Chroma EditUsed to customize chroma colors. To adjust the colors, use the soft menu’s dial-buttons [2], [3] and [4]. You ca
7-16Operation ManualColor MapSelect Color Map from the post curve menu. [Figure 7.8 Color Map] Color MapSelect a color map type. Use the Ext. Menu dia
7-17Chapter 7 Utilities D PostSelect D Post from the post curve menu. Configuration options are the same as with 2D Post. [Figure 7.9 D post]M PostS
7-18Operation Manual Preset This mode is used for system settings. It does not affect image output. The setup may be modified depending on specific ne
7-19Chapter 7 Utilities GeneralSelect the General tab on the Setup screen. From this tab, you can configure the general system settings.[Figure 7.11
7-20Operation Manual End Exam ActionUsed to assign a task to the control panel’s End Exam button. XEnd Exam Only: Pressing the End Exam button exit
7-21Chapter 7 Utilities Store Clip Setting Store Clip MethodSpecify the method and range in which an image is acquired and saved.You can select E
7-22Operation ManualDisplaySelect the Display tab in the Setup screen to specify display-related options. Display Option Select the items that you w
7-23Chapter 7 Utilities Font FontSpecify the target for which you want to set the font. Choose from Document Font and Measure Result Font. Font
6Operation ManualPreset ...
7-24Operation Manual TitleYou can specify the information that is displayed in the title area on the screen. InstituteEnter the name of the hospita
7-25Chapter 7 Utilities Date FormatUsed to configure the date format. Select a format by using the combo button. The date format that you specify
7-26Operation ManualAnnotateSelect the Annotate tab in the Setup screen. Specify display-related options.[Figure 7.14 Setup – Annotation] BodyMarker
7-27Chapter 7 Utilities BodyMarker Edit 1 2[Figure 7.15 BodyMarker Edit] 1 BodyMarker list: The list varies depending on the group selected from
7-28Operation ManualXSaving and Canceling the BodyMarker listClick Save to save the list. Click Close to cancel. XResetting the BodyMarker listClick
7-29Chapter 7 Utilities The system contains a list of abbreviated words for this function. You can add a new abbreviation or edit the existing abbre
7-30Operation Manual[Figure 7.17 Autotext Edit] Clear AnnotationCheck this checkbox to delete the entered annotation when you change the mode.
7-31Chapter 7 Utilities PeripheralsSelect the Peripherals tab on the Setup screen. You can configure keys, buttons and the peripheral devices connec
7-32Operation ManualNOTE: This is only supported by some digital printers. [Figure 7.18 Setup – Peripherals]
7-33Chapter 7 Utilities User Dened KeySelect the User Dened Key tab on the Setup screen. You can set the functions of the keys and buttons on the
7Table of ContentsChapter 9 ProbesProbes ...
7-34Operation Manual[Figure 7.19 Setup – User Defined Key]
7-35Chapter 7 Utilities MiscellaneousSelect the Miscellaneous tab on the Setup screen. You can set E-mail, Text, Network Status, etc.[Figure 7.20 Se
7-36Operation Manual Buzzer ControlGenerate a buzzer sound when a button or dial-button is used. Buzzer SoundSet this to On or O using the trackba
7-37Chapter 7 Utilities Network CongurationPress the Network Conguration button, and the TCP/IP Properties window will open to let you configure
7-38Operation ManualOptionSelect the Option tab on the Setup screen. Sets or releases the use of the optional software or hardware. [Figure 7.23 Setup
7-39Chapter 7 Utilities OptionsThe list of optional software will appear. NOTE: To purchase optional software, please contact the software’s distri
7-40Operation ManualDICOMSelect the DICOM tab on the Setup screen. Used to configure DICOM (Digital Imaging and Communication in Medicine) operation a
7-41Chapter 7 Utilities AE TitleEnter the name of the DICOM AE (Application Entity). The title is used to identify devices that use DICOM within
7-42Operation Manual Adding DICOM ServicesClick Add on the screen. A screen is displayed where you can enter a DICOM service to add. After adding a se
7-43Chapter 7 Utilities Retry IntervalSpecify how long the system will wait before it retries when transmission fails. You can specify this time
7-44Operation ManualStorage Server InformationSelect STORAGE under Services. Configure the Image Storage Service using DICOM. [Figure 7.25 DICOM Confi
7-45Chapter 7 Utilities VOI LUT SetupConfigure VOI LUT (Value Of Interest, Look Up Table). Adjust the brightness and contrast of a DICOM image w
7-46Operation ManualPrint Server InformationSelect PRINT under Services. Configure the Print Service using DICOM. NOTE: XYou can configure a printer
7-47Chapter 7 Utilities Medium TypeSpecify the material type for the print-out. Select from Paper, Clear Film, Blue Film, Mammo Clear Film and Ma
7-48Operation Manual[Figure 7.26 DICOM Configuration – Print]
7-49Chapter 7 Utilities Worklist Server InformationSelect WORKLIST under Services. Configure the Modality Worklist Service using DICOM. Show Work
7-50Operation ManualXPast Month: Retrieve the patient list for a month before the current date.XCustom Date: Specify a certain date and retrieve the
7-51Chapter 7 Utilities PPS Server InformationSelect PPS (Performed Procedure Step) under Services. Configure the Modality Performed Procedure Step
7-52Operation ManualStorage SR Server InformationSelect Storage SR (Storage Structured Report) under Services. Configure the Report Storage Service u
7-53Chapter 7 Utilities Managing DICOMClick Queue on the screen to switch to the DICOM Job Status screen. You can review the current job status usi
Cine/Loop ...6-3Annotation ...6-6Text
7-54Operation Manual7 Status: Displays the current status of the job. Status DescriptionFailThe job failed.TransferThe job is in progress.ImperfectJ
7-55Chapter 7 Utilities DeleteDeletes the selected job. This button is enabled only when the status of the selected job is Fail, Imperfect, Wait
7-56Operation Manual[Figure 7.32 DICOM Log]
7-57Chapter 7 Utilities AutoCalcSelect the AutoCalc tab on the Setup screen. AutoCalc is a Spectral Doppler mode feature that automatically performs
7-58Operation ManualAboutSelect the About tab in the Setup screen. Information about the system software version will be displayed. Click Detail to vi
7-59Chapter 7 Utilities Measurement Settings Select Measure Setup from the utility menu. Specify various setup options for measurements. The setup
7-60Operation ManualGeneralSelect the General tab on the Measure Setup screen. You can specify basic measurement options. GeneralSelect the sub-tab Ge
7-61Chapter 7 Utilities Apply Auto Resizing Line MarkerSelect whether to automatically reduce the marker cursor size when the distance between th
7-62Operation Manual DisplaySpecify items to display during measurement. Check the checkboxes of the items you wish to use. Show Measured Value in M
7-63Chapter 7 Utilities Measurement Results XTransparent BK color: Sets a transparent background. XNumber of Results Displayed: Specify the num
Comments to this Manuals