9-26
Operation Manual
5. Install the needle into the needle guide for use. Insert the needle until it reaches the examination
site.
To keep the needle securely in the needle guide, press down on the top of the biopsy adapter with
your index finger.
6. When the examination location is reached, take the needle out of the needle guide.
7. Detach the needle guide, adapter and sheath from the probe.
8. Discard all disposable items following use.
Needle Guide Alignment
Alignment of the needle guide displayed on the system is for the purpose of verifying whether the
needle and the needle guide are properly installed. This must be done prior to the biopsy examination.
If the needle does not follow the correct path when needle guide alignment is performed, contact
Samsung Medison’s service department for servicing.
Reverberation or other tissue artifacts may produce false needle images, which can cause confusion.
Ensure the needle path is along the guideline, and that you are not using a false needle image to
locate the needle.
WARNING:
X
The needle used for this alignment verification must not be used for the actual procedure.
Always use a new, sterile needle for each biopsy procedure.
X
To assist in accurate projection of the needle, use a straight, new needle for each alignment
procedure.
1. Attach the biopsy kit.
2. Set the system depth for the procedure to be performed and select the Biopsy menu.
3. Immerse the probe into the water bath, and insert the needle into the needle guide.
4. Conrm that the needle image is on the needle guidelines. If so, the needle guide is properly aligned.
5. If the needle image is out of the needle guideline, check the needle guide or the probe adapter.