5-5
Chapter 5 Measurements and Calculations
Optimization of Measurement Accuracy
2D Mode
X
Resolution is in proportion to the frequency of the probe.
X
Penetration is in inverse proportion to the frequency of the probe.
X
The highest resolution can be obtained at the focus of the probe where the ultrasound beam is
narrowest.
X
The most accurate measurements can be obtained at the focus depth. The accuracy decreases as
the distance from the focus increases, widening the beam width.
X
Using the zoom function or minimizing the depth display makes distance or area measurements
more accurate.
M Mode
X
The accuracy of time measurements can be increased when the sweep velocity and the display
format are set to high values.
X
The accuracy of distance measurements can be increased when the display format is set to high
values.
Doppler Mode
X
It is recommended to use lower frequency ultrasound for measurement of faster blood ows.
X
The size of the sample volume is limited by the axial direction of the ultrasound.
X
Using lower frequency ultrasound increases penetration.
X
The accuracy of time measurements can be increased when the sweep velocity is increased.
X
The accuracy of velocity measurements can be increased when the vertical scale is set to smaller
values.
X
It is most important to use an optimal Doppler angle to enhance the accuracy of velocity
measurements.