Acoustic Measurement Precision and Uncertainty
The Acoustic Measurement Precision and Acoustic Measurement Uncertainty are described below.
Quantity Precision Total Uncertainty
Ipi,α (attenuated pulse intensity integral)
3.2% +21% to -24%
P (acoustic power)
6.2% +/- 19%
P
r,α
(attenuated rarefaction pressure)
5.4% +/- 15%
f
awf
(acoustic working frequency)
< 1% +/- 4.5%
Systematic Uncertainties
For the pulse intensity integral, de-rated rarefaction pressure (Pr.3), center frequency, and pulse
duration, the analysis includes considerations of the eects on accuracy of:
Hydrophone calibration drift or errors.
Hydrophone / Amp frequency response.
Spatial averaging.
Alignment errors.
Voltage measurement accuracy, including.
X
Oscilloscope vertical accuracy.
X
Oscilloscope oset accuracy.
X
Oscilloscope clock accuracy.
X
Oscilloscope Digitization rates.
X
Noise.
The acoustic power is measured using a Radiation Force for systematic uncertainties through the
use of calibrated NIST acoustic power sources.
We also refer to a September 1993 analysis conducted by the working group of the IEC technical
committee 87 and prepared by K. Beissner, as a rst supplement to IEC publication 1161.
The document includes analysis and discussion of the sources of error/measurement eects due
to:
X
Balance system calibration.
X
Absorbing (or reecting) target suspension mechanisms.