Chapter 1 Safety
Scale
Using the scale control to increase the color velocity range may increase the TI. The system will
automatically adjust the pulse voltage to stay below the system maximum. A decrease in pulse
voltage will also decrease the MI.
2D Mode Size
A narrower 2D-mode sector width in Color imaging will increase the color frame rate. The TI will
increase. The MI will not change. If pulsed Doppler is also enabled, then pulsed Doppler will
remain as the primary mode and the TI change will be small.
M Mode and Doppler Controls
Simultaneous and Update Methods
Use of combination modes aects both the TI and MI through the combination of pulse types.
During Simultaneous mode, the TI is an ancillary element. During Auto-update and Duplex, the
TI will display the dominant pulse type. The displayed MI will be from the mode with the largest
peak pressure.
Sample Volume Depth
When the Doppler sample volume depth is increased, the Doppler PRF may automatically
decrease. A decrease in PRF will decrease the TI. The system may also decrease the pulse voltage
to remain below the system maximum. A decrease in pulse voltage will decrease the MI.
Other
2D, Color, M, Doppler Modes (2D, Color, M mode, PW)
When a new imaging mode is selected, both the TI and the MI will revert to their default settings.
Each mode has a corresponding pulse repetition frequency and maximum intensity point. In
combined or simultaneous modes, the TI is the sum of the contribution from the modes enabled,
and the MI is the value for the focal zone of the mode with the largest de-rated intensity. If a mode
is turned o and then reselected, the system will return to the previously selected settings.