Operation Manual
Entering & Exiting PW Spectral Doppler Mode
Tap the PW button on the touch screen. Tap the Update button on the touch screen to obtain a
spectral Doppler image.
NOTE: The Doppler image can only be obtained in the D Only or Simultaneous states.
Sample Volume for PW Spectral Doppler Mode Screen
The Doppler spectrum is displayed when the Sample Volume is located above the blood ow on
the 2D image. The size of the Sample Volume is displayed in mm units. Select the sample volume
with your nger to adjust its position.
X
Moving Sample Volume: Select the sample volume of the study image to adjust its position.
X
Adjusting Sample Volume Angle: Tap the Angle button’s ‘+/-‘ buttons on the touch screen to
select an angle between -80 and 80 degrees. Alternatively, tap the Quick Angle button on the
touch screen to select a preset angle of -60, 0, or 60 degrees.
Adjusting Doppler Baseline
Tap the Baseline ‘+/-‘ buttons on the touch screen to adjust the Doppler baseline.
PW Spectral Doppler Mode Menu
Invert
Tap Invert on the touch screen. Tapping the button inverts ‘+ and -’, which indicate the velocity of
spectrum.
Simul. (Simultaneous)
Tap Simul. on the touch screen. Each time you tap the button, the Simultaneous function turns
on or o. If the Simultaneous option is turned on, you can view real-time 2D and spectral Doppler
images at the same time. If the option is not turned on, however, you will only be able to view
the image from one of these modes. The Simultaneous function decreases Doppler PRF, thus
decreasing the measurable speed range.