9-17
Chapter 9 Probes
Cleaning
Cleaning is an important procedure that is carried out before disinfecting the probe. The probe must
be cleaned after each use.
CAUTION:
X
Do not use a surgical brush to clean probes. Even the use of soft brushes can damage the probe.
X
During cleaning and disinfection, keep the parts of the probe that must remain dry higher than
the other parts during wetting, until all parts are dry.
1. Disconnect the probe from the system.
2. Remove any biopsy adapters or needle guides. (Adapter may be reused after disinfection.)
3. Remove the sheath. (Sheaths are disposable.)
4. Use a soft cloth, lightly dampened with mild soap or compatible cleaning solution, to remove any
particulate matter and bodily uids that remain on the probe or cable.
5. To remove remaining contaminants, rinse with water up to the immersion point.
6. Wipe with a dry cloth.
7. If necessary, wipe rst with a water-dampened cloth to remove soap residue, and wipe with a dry
cloth again.
Comments to this Manuals