Samsung PT60A User Manual Page 122

  • Download
  • Add to my manuals
  • Print
  • Page
    / 228
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 121
5-6
Operation Manual
Color/Power Doppler Mode
X
A protocol is not specied for images in Color Doppler Mode or Power Doppler Mode.
Therefore, the same limitations imposed when measurements are taken in B/W images apply to
the accuracy of the measurements taken in these modes.
X
Using Color/Power Doppler Mode images to measure accurate blood ow velocity is not
recommended.
X
The amount of blood ow is calculated based on the average velocity, rather than the peak
velocity.
X
In all applications, the amount of blood ow is measured in PW Spectral Doppler Mode.
Cursor Position
X
All measurements are aected by input data.
X
Adjust the images on the screen so that they are displayed at maximum granularity.
X
Use the front edge or boundary point of a probe to make the start and end points of a
measurement object more distinct.
Make sure that the probe direction remains aligned during measurement.
Page view 121
1 2 ... 117 118 119 120 121 122 123 124 125 126 127 ... 227 228

Comments to this Manuals

No comments