9-23
Chapter 9 Probes
Cleaning and Disinfecting the Biopsy Kit
WARNING: Always use protective eyewear and gloves when cleaning and disinfecting biopsy kit.
Wash and disinfect the biopsy kit to reduce pathogens to the level of 10
-6
. Some components of the
biopsy kit may be disposable. Please read the biopsy kit user manual carefully before use.
Cleaning and Disinfection of Stainless Steel Biopsy Kit
Cleaning
1. After use, remove the biopsy kit from the probe.
2. Disassemble the biopsy kit into its component parts, if applicable.
3. Using a small brush and water, scrub each part to remove trapped material from the biopsy kit.
4. Rinse with water to remove remaining particulates.
Disinfection
1. Disinfect the adapter by autoclaving (Steam) or using gas (Ethylene Oxide).
2. After disinfection, follow the proper post-disinfection procedure for the disinfection method
used (refer to the instructions for the disinfecting equipment).
3. Inspect the components for corrosion, breakage, and other damages. If damage is discovered,
please contact the Samsung Medison customer support department.